HL / SL
49
9.5. Control
The circulator is designed for internal and external temperature control.
This can be switched in this submenu. Only the set of parameters
corresponding to the particular setting is displayed.
Control internal Control external
For external temperature control and measurement connect a Pt100
external sensor to the socket at the rear of the circulator.
Press the key if a parameter is to be retained. (prior to OK)
Level 2 Parameter-Level
or
Switch over control type by pressing
and
Possible parameters: intern / extern
The control type can be adjusted in the
-OFF- mode only.
The parameter flashes, switch by pressing
and
Possible parameters: off / once / always.
Control Type internal Parameter-Level
The parameter flashes, switch by pressing
and
Possible parameters: aperio / normal
This parameter affects the temperature
sequence in case of internal control.