54
Optimization instructions for the PID control parameters:
optimum setting
°C
t
The heat-up curve reveals inappropriate control
settings.
Inappropriate settings may produce the following heat-up curves:
Xp too low
°C
t
Tv/Tn too low
°C
t
Xp too high or Tv too high
°C
t
Tv/Tn too high or Xp too high
°C
t