Safety installations, warning functions
28
O
verTemp 87.0°C
S
ubTemp 83.0°C
Setpoint3 85.0°C
The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within
the set limits for 3 seconds.
Setting >WARNING<
A mere warning function with optical and audible warning signal
(interval tone) A message appears on the VFD COMFORT-
DISPLAY:
or
OverTemp SubTemp
The LCD –DISPLAY shows the message:
Setting >ALARM<
Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message appears
on the VFD COMFORT-DISPLAY:
or
OverTemp SubTemp
The LCD -DISPLAY shows the message
Press
for help on the LCD

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Julabo SL-12 and is the answer not in the manual?

Julabo SL-12 Specifications