EasyManua.ls Logo

Juvent 1000 - Test Operation

Juvent 1000
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Juvent 1000 Instruction Manual
Test Operation
11. Stand on the device and place your special Juvent test card over the “Start
Treatment” area and verify the following:
A. The red RFID standby indicator went off and the green treatment
indicator turned on.
B. The device’s blue LCD lights up.
C. The device runs through the 12 second “Self test cycle” and begins the
10 minute treatment countdown.
If your first card did not activate the system, repeat the test operation with the second
card.
12. If the system still does not activate, check all connections.
13. Troubleshooting: See page 24 or call Juvent engineering - 1-877-209-9600
NOTE: Periodically ensure that all electrical connections are properly in place.