Questions and Answers

  • C
    Courtney AyersAug 8, 2025
    What to do if the Kayden Switch shows a Self-Test Error?
    • D
      Dana DavisAug 8, 2025
      If your Kayden Switch displays a Self-Test Error, the issue could stem from the Electronics Module, internal connections, external wiring, or the sensor. First, disconnect the Electronics Module and check the sensor wiring as shown in Figure 24. If the wiring resistance values are correct, the fault may be in the Electronics Module, which should be returned to the factory for testing and repair. If the wiring resistance values are incorrect, the sensor should be returned for testing and repair.