8. Measurement procedure
8.2 Preparing the patient
Use an appropriate topical anaesthetic to numb the cornea.
Place a strip of fluorescein-soaked paper near the outer
canthus in the lower conjunctival sac. After a few seconds the
lacrimal fluid will be coloured and the paper may be removed.
When using drops, we recommend a 0.5% solution of
fluorescein sodium. If using a 1% or 2% solution, use a glass
rod to introduce a small quantity of liquid into the
conjunctival sac.
Seat the patient at the Slit Lamp and place his / her chin on
the chinrest and ensure the forehead touches the forehead rest.
Adjust the chinrest height so that the patients eye is at the
correct height (most Slit lamps have a marker on the chinrest
pole for correct height adjustment).
Reusable prisms are not disinfected prior to shipping and must
be cleaned before their first use as described in section 4 of this
document.
Check prisms for damage prior to use.