8. Measurement procedure
8.4 Using the instrument / taking a measurement
8.4.1 Instructions to the patient
The patient’s head must be firmly positioned on the chin rest
and the forehead rest. If necessary, a band may be used to
hold the head still.
Ask the patient to look straight ahead. If necessary, use a
fixation target to keep the eyes still.
We recommend occasionally reminding the patient during the
examination, to keep his / her eyes wide open. If necessary,
the examiner may use the tips of his fingers to hold the lids
open, taking care not to exert pressure on the eye.
When elevating the lids, the angle between the microscope
and the lighting unit must be reduced to about 10° so
that the light beam passes through the body of the prism. In
this position it should be possible to obtain an image with no
reflections.
Immediately before measurement, ask the patient to close
his / her eyes for a few seconds, in order to ensure that the
cornea be sufficiently wetted by the lacrimal fluid containing
the fluorescein solution.