9. Problem solving
9.2 The measurement prism does not touch the cornea or too much force has been applied
Home
Back
Next
Cause:
If the patient pulls his/her head back even
slightly, the pulses will become irregular and
the measuring prism contact with the eye
will become intermittent. If the patient pulls
his / her head even further back, the
fluorescein semi-circles will completely
disappear.
Correction:
If possible, use a band to hold the patient’s
head in place.
Cause:
If during measurement the slit lamp is moved
forward toward the patient or the patient moves
toward the slit lamp, the sensor arm will be pushed
into contact with a stop spring and an audible
alarm will sound. The applanation surface will be
too large. The image will not change when the
measurement drum is rotated.
Correction:
Retract the slit lamp until regular pulses and a
correspondingly smaller applanation surface
are obtained. This is the correct measurement
position, in which variations in pressure
will not cause immediate variations in
the applanation surface.