11. Annex I – EMC statement and guidelines
Guidance and manufacturer’s declaration – electromagnetic immunity
The Keeler Vantage Plus LED is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in
such an environment.
Electromagnetic environment - guidance
Electrostatic
discharge (ESD).
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst.
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines
N/A
Mains power quality should be that of a typical commercial or
hospital environment.
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
N/A
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
<5% UT
(> 95% dip in UT) for 0.5 cycles
40% UT
(60% dip in UT ) for 5 cycles
70% UT
(30% dip in UT ) for 25 cycles
<5% UT
(>95% dip in UT) for 5 s
<5% UT
(> 95% dip in UT) for 0.5 cycles
40% UT
(60% dip in UT) for 5 cycles
70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT ) for 5 s
Mains power quality should be that of a typical commercial or
hospital environment.
If the user of the Keeler Vantage Plus LED requires continued
operation during power mains interruptions, it is recommended
that the charger be powered from an uninterruptible power
supply.
Power frequency
(50/60 Hz) magnetic
field. IEC 61000-4-8
Power frequency magnetic fields should be at a level
characteristic of a typical location in a typical commercial or
hospital environment.
Note UT is the a.c. mains voltage prior to application of the test level.