31
11. Annex I EMC statement and guidelines
Guidance and manufacturer’s declaration – electromagnetic emissions
The Keeler Vantage Plus LED is intended for use in the electromagnetic environment specified below. The customer or user should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Charger only
RF emissions
CISPR 11
Group 1
The Keeler Vantage Plus LED uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Keeler Vantage Plus LED is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Indirect
Ophthalmoscope only
RF emissions
CISPR 15
Complies
The Keeler Vantage Plus LED is not suitable for interconnection with other
equipment