EasyManua.ls Logo

Keithley 2001-TCSCAN - Table; Verification and Calibration Equipment

Keithley 2001-TCSCAN
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Information
4-3
V
REF
= (T
J
+ 273.15)
×
0.0002
7. Read the voltage across the output of OUT A (Model
2001 #2). Compare the measured and calculated voltag-
es. If they differ by more than 72
µ
V (0.36
°
C), perform
the calibration procedure of paragraph 4.4.1.
Table 4-1
Verification and calibration equipment
Description Model or part Specifications Applications
DMM* Keithley Model 2001 20, 72ppm
200k, 90ppm
200mV, 37ppm
Path resistance
Calibration
Reference junction, calibration
Electrometer w/voltage source Keithley Model 6517 20pA, 200pA; 1.6%
200V source; 0.05%
Offset current, path isolation
Sensitive Digital Voltmeter Keithley Model 182 3mV; 60ppm Contact potential
Triax cable (unterminated) Keithley Model 237-
ALG-2 Offset current
Low thermal cable (untermi-
nated)
Keithley Model 1484 Contact potential
Thermistor probe Thermometrics Series
CSP A207A
±0.005°C accuracy Reference junction, calibration
Reference thermometer ±0.01°C w/thermistor Reference junction, calibration
Distilled water ice bath
(Dewar flask or Thermos)
±0.1°C Calibration
#22 AWG solid copper wire Path resistance
#22 AWG Teflon-insulated
stranded wire
Path isolation
* Two DMMs are required for thermistor calibration method and reference junction test.
4. Coat the probe with a thermally conductive compound,
and insert it into the 0.110” hole marked “CAL” on the
Model 2001-TCSCAN printed circuit board.
5. Use the Model 2001 #1 front panel CHAN key to close
Channel 1.
6. Take a reading from the reference thermometer (T
J
). Use
the following equation to calculate the equivalent refer-
ence junction output voltage (V
REF
):

Table of Contents

Other manuals for Keithley 2001-TCSCAN

Related product manuals