Specications
Table 7: Common characteristics (cont.)
Item Description
Isolation Voltage, O utput to
Chassis
Any output can be oated up to 240 V (DC + peak AC with AC limited to a maximum 3 V
pk-pk
and
maximum of 60 Hz) relative to the Earth Ground terminal. Note that remote sense terminals
should remain within 1 V of their respective outputs to avoid possible damage.
Isolation Voltage, Output to Output Any output can be oated up to 240 V (DC + pk AC with AC limited to a maximum of 3 V
pk-pk
and a maximum of 60 Hz) relative to any other output terminal. Remote sense terminals should
remain within 1 V of their respective outputs to avoid possible damage.
Table 8: Terminal strip
Item Description
The rear panel has a customer accessible terminal strip. This strip is comprised of a board-mounted section and a
technician-removeable section. The technician-removeable section offers the customer a safe set o f screw terminals to connect
to vario
us instrument functions detailed below.
Table 9: Interfaces and output ports
Parameter All m odels
USB device,
nominal
USB 2.0 Full Speed device. 12 Mb/sec maximum.
Command
processing time,
typical
About 400 ms per command.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 9

Other manuals for Keithley 2220-30-1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2220-30-1 and is the answer not in the manual?

Keithley 2220-30-1 Specifications

General IconGeneral
BrandKeithley
Model2220-30-1
CategoryPower Supply
LanguageEnglish

Related product manuals