EasyManua.ls Logo

Keithley 2220-30-1 - Page 52

Keithley 2220-30-1
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Ve rication
Check DC Current Line
Regulation
This check proc
edure uses the same test setup as the previous procedure.(See
Figure 6.)
For each channel sequentially.
1. Change the AC Power Source output to the voltage specied in the following
table.
DUT voltage selector
switch AC power source voltage Japan congured units
110 99 V 90 V
220 198 V 180 V
NOTE. Ensure the warm-up criteria has been met as described in the Performance
Verication Conditions.
2. Set the voltmeter as follows:
a. Set to measure DC volts.
b. Set to auto range.
c. Set to show amps (instead of volts) by m ultiplying the voltmeter result
b
y20.
Press Shift DC V (mX+B).
Use the arrow key to move the cursor to the far right until it is
positioned on the ^.
Press the up range button once so that M=10.00000.
Press the arrow key to move the cursor to the 1’s digit and then press
the up range button once so that M=20.00000.
Press the Enter button and check that the display shows B=0.
Press the Enter button and check that the display shows UNITS.
Use the arrow keys to modify the display to show AMP.
Adjust each letter individually.
3. Set the electronic load as follows:
a. Set to Constant Voltage.
b. Set to the specied voltage for y our channel under test (CUT). (See
Table 22 on page 22.)
4. Set the channel under test (CUT) to 100% of the full scale (FS) output voltage.
5. Set the CUT to 100% of the full scale (FS) output current.
42 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference

Other manuals for Keithley 2220-30-1