2260B Series User Manual
44
Before setting the OVP or OCP level:
Ensure the load is not connected.
Ensure the output is set to off.
Steps
1. Press the OVP/OCP key. The
OVP/OCP key lights up.
2. The OVP setting will be displayed on the top
and the OCP setting (or OFF) will be displayed
on the bottom.
V
A
OVP Setting
OCP Setting
OVP Level
3. Use the Voltage knob to set the
OVP level.
Voltage
Range
10%~110% of rated
output voltage.
OCP Level
4. Use the Current knob to set the
OCP level.
Current
Range
10%~110% of rated
output current, OFF.
5. Press OVP/OCP again to exit. The
OVP/OCP indicator will turn off.
Power switch trip
Set F-95 (Power switch trip) to 1 (to
disable the power switch trip) or to
0 (to enable the power switch trip)
and save.
Page 96
F-95
1 (Disable) or 0 (Enable)