REMOTE CONTROL 
57 
[SOURce:]CURRent:PROTection:STATe 
 
Description 
Turns OCP (over-current protection) on or off.  
Syntax [SOURce:]CURRent:PROTection:STATe {0|1|OFF|ON} 
Query Syntax  [SOURce:]CURRent:PROTection:STATe? 
Parameter/Return
0 
<NR1> Turns the buzzer off. 
  OFF 
Turns the buzzer off. 
  1 
<NR1> Turns the buzzer on. 
  ON 
Turns the buzzer on. 
Return parameter
<Bool> 
Returns bleeder resistor status (0 or 1). 
Example SOUR:CURR:PROT:STAT OFF 
Turns OCP off. 
[SOURce:]CURRent:SLEW:RISing 
 
Description 
Sets or queries the rising current slew rate. This is 
only applicable for CC slew rate priority mode. 
Syntax 
Query Syntax 
[SOURce:]CURRent:SLEW:RISing {<NRf>|MIN|MAX} 
[SOURce:]CURRent:SLEW:RISing? [MIN|MAX] 
Parameter/Return
<NRf> 
0.01A/s~200% (2260B-30-36) 
0.1A/s~200% (2260B-30-72) 
0.01A/s~200% (2260B-80-13) 
0.01A/s~200% (2260B-80-27) 
  MIN 
Minimum rising current slew rate. 
  MAX 
Maximum rising current slew rate. 
Example SOUR:CURR:SLEW:RIS 72 
Sets the rising current slew rate to 72A/s. 
[SOURce:]CURRent:SLEW:FALLing 
 
Description 
Sets or queries the falling current slew rate. This is 
only applicable for CC slew rate priority mode. 
Syntax [SOURce:]CURRent:SLEW:FALLing {<NRf>|MIN|MAX} 
Set
Query
Set
Query
Set
Query