2260B Series Programming Manual
90
Error Code Description
-100 Command
Error
This is the generic syntax error for devices that
cannot detect more specific errors. This code
indicates only that a Command Error as
defined in IEEE 488.2,11.5.1.1.4 has occurred.
-102 Syntax error
An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.
-103 Invalid
separator
The parser was expecting a separator and
encountered an illegal character; for example,
the semicolon was omitted after a program
message unit,
MEAS:VOLT:DC?:MEASCURR:DC?
-104 Data type
error
The parser recognized a data element different
than one allowed; for example, numeric or
string data was expected but block data was
encountered.
-108 Parameter
not allowed
More parameters were received than expected
for the header; for example, the KLOCk
command only accepts one parameter, so
receiving SYSTem:KLOCk 1,0 is not allowed.
-109 Missing
parameter
Fewer parameters were received than required
for the header; for example, the KLOCk
command requires one parameter, so receiving
KLOCk is not allowed.
-111 Header
separator error
A character which is not a legal header
separator was encountered while parsing the
header; for example, no blank space followed
the header, thus
APPL 5, 1 is an error.

Table of Contents

Other manuals for Keithley 2260B-30-72

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-30-72 and is the answer not in the manual?

Keithley 2260B-30-72 Specifications

General IconGeneral
BrandKeithley
Model2260B-30-72
CategoryPower Supply
LanguageEnglish

Related product manuals