Section
8: Constant voltage adjustment Series 2260B Programmable Power Supplies
Verification and Adjustment Manual
8-4 077104802 / November 2016
The "x" in the diagram above indicates that the value is not fixed.
12. Rotate the current knob to set the middle voltage value on the 2260B display so that it is the
same as the reading on the PCS-1000 voltmeter.
13. Press the voltage knob to enter the output voltage full-scale adjustment (C115).
14. Press the voltage knob to set the full-scale voltage value (C116).
The "x" in the diagram above indicates that the value is not fixed.
15. Rotate the current knob to set the full-scale voltage value on the 2260B display so that it is the
same as the reading on the PCS-1000 voltmeter.
16. Press the voltage knob to enter the OVP (overvoltage protection) adjustment (C117).
17. Press the voltage knob to start the automatic OVP adjustment (C118).