8: Constant voltage adjustment Series 2260B Programmable Power Supplies
Verification and Adjustment Manual
8-6 077104802 / November 2016
22. Press the voltage knob to exit adjustment mode.
The "x" in the diagram above indicates that the value is not fixed.
Constant voltage adjustment is complete.