APPENDIXF
236/237
Performance
Verification
Table
F-5.
Voltage
Measure
Verification
236/237
Compliance
Limit
Allowable
Reading
343ARange
343AOutput
&
Measure
Range*
(18°C
to
28°C)
10V
ov
1.1V
10V
ov
11V
10V
ov
110V
10V
ov
1100V
10V
+1V
1.1V
10V
-1V
1.1V
10V
+10V
11V
10V -10V 11V
100V
+100V
110V
100V
-100V
110V
1000V
+1000V 1100V
1000V
-1000V 1100V
*236/237
set to source
OnA
on
lnA
source range.
3. Temporarily disconnect
the
banana
cable from
VOLTS
ill
of
the
DMM
at
the
rear
panel
of
the test
fixture (BINDING
POST
1),
and
connect
itto
VOLTS
LO
of
the
DMM. This shorts the
input
of
the
Model
193ADMM.
4.
On
the
Model 193A DMM, enable
ZERO
to
cancel
any
offsets.
+
-0.00030V
to
+0.00030V
-00.0010V
to
+00.0010V
-000.010V
to
+000.010V
-0000.10V
to
+0000.10V
+0.99942V
to
+1.00058V
-0.99942V
to
-1.00058V
+9.996SV to
+10.0035V
-9.9965V to
-10.0035V
+99.965V to
+100.035V
-99.96SV to
-100.035V
+999.55V
to
+1000.45V
-999.55V
to
-1000.4SV
5. Re-connect
the
banana
cable
to
BINDING
POST
1 of
the
test fixture.
6. Place
the
Model236/237
in
operate
and
verify
that
the
reading
on
the
DMM
is
within
the
following al-
lowable limit;
9.99500 V
to
10.00500V.
7.
Place the
Model236/237
in
standby
and
local sense.