2600AS-900-01 Rev. B / September 2008 Return to Section Topics 3-11
Series 2600A System SourceMeter
®
 Instruments User’s Manual Section 3: Test Script Processor Interaction
Figure 3-5
Run configuration example - Script Attributes tab shown
Running the user script
Remote programming
The functions for the KIGeneralUser user script can be called after the script is run as follows:
KIGeneralUser()
After the above command is executed, any of the functions that make up the user script can be 
called in the same manner that factory script functions are called (see 
Step 1: Call the function).
Front panel operation
To run the script from the front panel, a test name has to be added to the USER TESTS submenu 
of the LOAD TEST menu. Since the KIGeneralUser script is made up of eight functions, a front 
panel user test must run the script and also call a function. The following command will add the 
test name “PulseVMeasI” to the menu. It will also configure the test to run the 
KIGeneralUser 
script and call the 
PulseVMeasureI function: 
display.loadmenu.add("PulseVMeasI", "KIGeneralUser() PulseVMeasureI()", 
display.SAVE)
Because no parameters are used in the function (PulseVMeasureI()), the test will be interactive 
and require the operator to input the test parameters using the front panel controls. See 
How do I 
run a script from the front panel? for details on running a user script from the front panel.
The command display.loadmenu.add can be used to configure front panel user tests for the 
other functions of the script. Make sure to use a different menu name for each function. For