EasyManuals Logo
Home>Keithley>Multimeter>3700A Series

Keithley 3700A Series Firmware Release Notes

Keithley 3700A Series
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Series 3700A Firmware Release Notes Version 1.52 Release - Non-critical fixes
3706A-850SV1.55a / June 2017
15
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When communicating with the product using USB and using the Read Status Byte bus operation it
always returned 0.
Workaround:
None.
Resolution:
It now returns the correct status byte.
PR47137

Other manuals for Keithley 3700A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700A Series and is the answer not in the manual?

Keithley 3700A Series Specifications

General IconGeneral
BrandKeithley
Model3700A Series
CategoryMultimeter
LanguageEnglish

Related product manuals