FAQs
How should I clean the front-panel display?
If you need to clean the front-panel LCD touchscreen display,
use a soft dry cloth. If necessary, use a cloth dampened with an
ammonia-free glass cleaner. Do not spray cleaning uids onto the
display. Additionally, it is strongly recommended that you use only
ngers on the touchscreen. Do not use sharp objects, such as a
pen or pencil, to touch the touchscreen.
My data looks odd or is wrong. What should I do?
Verify the connections from the instrument to the test xture. Also
check the connections from the DUT to the test xture socket.
I cannot unplug the mini-triaxial cable (4200A-MTRX) from
the SMU. What should I do?
The mini-triaxial connector is a locking connector. To remove it,
pull the knurled part of the connector back.
A con guration error has been detected and I cannot
launch Clarius. What should I do?
This occurs when the physical con guration does not
match the KCon de ned con guration or when there are
communication problems between instruments and the
4200A-SCS.
To verify system con guration:
1. Run KCon.
2. Select Validate Con guration.
This can also occur if the preampli er or RPM is removed or
reconnected.
To update the system con guration:
1. Run KCon.
2. Select Upgrade Preamp, RPM, and CVIV Con guration.
Note that preampli ers are SMU speci c. For example, a
preampli er that is con gured for SMU1 cannot be connected
to SMU2.

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals