DISPLAY MESSAGES
The following massages will be displayed to flag errors or conditions as
they occur:
MESSAGE
OL
r.r.
n Err
b Err
t EN
. .
DESCRIPTION
Overrange input applied.
Displayed briefly on power up;
RAM test passed. Displayed con-
stantly; RAM test failed.
Number Error. Calibration or
voltage source value exceeds
limits.
Bus Error. Instrument pro-
grammed while not in remote or
an illegal command (IDDC) sent or
an illegal command option
UDDCO) sent.
Trigger Overrun Error. Instrument
triggered while processing reading
from previous trigger.
flashing decimal points on expo-
nent LEDs; not using stored cal
constants. Constant decimal
points; instrument in calibration
male
IEEE “address” Program message; in IEEE ad-
dress program.
dlSP Program message; in display
mode program.
CAL
Program massage; in cal constant
program.
31inking AMPS light
In V/I current overload.
3linking ZERO CHECK light In V/I zero check.