S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual Section 3:
LPTLib command reference
S530-900-01 Rev. E / September 2017 3-43
Details
The lorangeX command is used with autoranging to limit the number of range changes, which
saves test time.
If the instrument is on a range lower than the one specified by the lorangeX command, the range is
changed. The S530/S540 automatically provides any settling delay for the range change that may be
necessary due to this potential range change.
When the force and measure functions are both current (I) or both voltage (V), specifying a lower
force range overrides the lowrangeX command.
Once defined, the lorangeX command is in effect until a devclr or devint command, or another
lorangeX command executes.