S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual Section 3:
LPTLib command reference
S530-900-01 Rev. E / September 2017 3-61
Example
double icer2;
.
.
conpin(3, 2, GND, 0);
conpin(SMU1, 4, 0);
limiti(SMU1, 1.0E-3); /* Limit current to 1.0 mA. */
rangei(SMU1, 2.0E-3); /* Select range for 2 mA. */
forcev(SMU1, 35.0); /* Force 35 V. */
measi(SMU1, &icer2); /* Measure leakage; return */
/* results to icer2. */
This example specifies connections, sets a 1 mA limit on the 2 mA range and forces 35 V, then measures
current leakage and returns the results to the variable icer2.
Equation 8: SMU measure and source range function
Also see
None