Multifuge
®
1 S / 1 S-R
Instruction Manual

Table of Contents

Questions and Answers

  • L
    Lisa OneillSep 18, 2025
    What does "Lift Lid" mean on my Kendro Heraeus Multifuge 1 S?
    • M
      Melissa WilliamsonSep 18, 2025
      If the message "Lift Lid" appears in the display of your Kendro Laboratory Equipment, it means that the lid has not been unlocked after release. Avoid placing objects onto the centrifuge lid and lift the lid slightly.
  • M
    martinfeliciaAug 25, 2025
    What to do if the lid of my Kendro Laboratory Equipment won't open?
    • M
      Mark WebbAug 25, 2025
      If you are unable to open the lid of your Kendro Laboratory Equipment, first ensure that the main power connection is working and the instrument is switched on, verifying that the display is lit. Next, press the lid down firmly in the middle of the front section while simultaneously pressing the "open lid" key. If the lid still won't open, you can use the emergency lid release.
  • G
    Gina WoodwardAug 26, 2025
    What does "bAl" mean on my Kendro Heraeus Multifuge 1 S Laboratory Equipment display and how to fix it?
    • T
      Tammy MillerAug 26, 2025
      If the message "bAl" appears on the display of your Kendro Laboratory Equipment, it indicates that the imbalance switch has been activated. To resolve this, open the instrument by pressing the "open lid" key. Check that the rotor is properly loaded and inspect for any broken tubes or damage to the rotor that may have triggered the switch. Also, ensure that the trunnions of the swinging bucket rotor are adequately lubricated.
  • S
    Sarah WalkerAug 27, 2025
    How to fix dark displays on my Kendro Laboratory Equipment?
    • K
      kim71Aug 27, 2025
      If the displays on your Kendro Laboratory Equipment remain dark, it indicates a mains voltage disconnection. First, verify that the mains plug is turned on and check the main power connection. If the main connection is confirmed to be working, it may indicate a more complex issue.
  • C
    colemanruthSep 22, 2025
    Why does my Kendro Laboratory Equipment display "OPEN" when the lid is closed?
    • M
      matthewbakerSep 22, 2025
      If the display shows "OPEN" even when the lid is closed on your Kendro Laboratory Equipment, it indicates that the lid is not properly closed. Open the lid and repeat the locking procedure.
  • K
    Kyle BaldwinSep 23, 2025
    Why do the displays on my Kendro Laboratory Equipment fail briefly?
    • E
      Eddie WilsonSep 23, 2025
      If the displays on your Kendro Laboratory Equipment fail briefly, it is because the main connection was briefly interrupted. Turn off the main switch, check whether the mains power cord is connected properly, and then restart the centrifuge.
  • M
    Michael JohnsonSep 9, 2025
    What does error E-28 mean on my Kendro Heraeus Multifuge 1 S?
    • B
      brettharrisSep 9, 2025
      If error code E-28 appears on your Kendro Laboratory Equipment, it indicates that the motor is running in the wrong rotational direction. Switch the instrument off and on again.
  • D
    deniseduncanSep 12, 2025
    What does error E-30 mean on my Kendro Heraeus Multifuge 1 S Laboratory Equipment?
    • M
      Maria BrownSep 12, 2025
      If error code E-30 appears on your Kendro Laboratory Equipment, it indicates that the control voltage breaks down. Switch the instrument off and on again.
  • C
    Christopher TaylorSep 3, 2025
    What does error E-33 mean on my Kendro Heraeus Multifuge 1 S?
    • M
      Michael CrawfordSep 3, 2025
      If error code E-33 appears on your Kendro Laboratory Equipment, it indicates overpressure in the refrigeration system. Turn the instrument off and pull the main plug. Check and clean the ventilation slots if necessary. After about 60 minutes, you can restart the instrument, ensuring you observe the maximum permissible environmental temperature!
  • D
    Donna BatesSep 1, 2025
    What does error E-31 mean on my Kendro Laboratory Equipment?
    • K
      kirbyanaSep 1, 2025
      If error code E-31 appears on your Kendro Laboratory Equipment, it indicates overtemperature in the motor. Turn the instrument off and unplug the mains power cord. Check and clean the ventilation slots if necessary. After about 60 minutes, you can restart the instrument, ensuring you observe the maximum permissible environmental temperature.

Kendro Heraeus Multifuge 1 S Specifications

General IconGeneral
Max Speed15, 000 RPM
Maximum Speed15, 000 RPM
Capacity4 x 250 ml
Temperature RangeNot applicable
Timer0-99 minutes
Noise Level<65 dB
Power Requirements230 V, 50/60 Hz