6. Product Main Features
7. Product Requirements and
Major Parametric Description
8. Method of application
7.1 Product Power Requirement
a) Power source: Rechargeable lithium battery3.7v/100mAh
b) Device safety classes: class II type BF
7.2 Product Main Technical Parametric description
a)Impulse frequency:
TENS:20-100Hz
b)Pulse width:
TENS: 120µs
c)Impulse waveform: square wave
d)The product output has no the DC component.
e)Influence of output end open circuit and short circuit: it is able to support
the influence of output end open and short circuit, and its performance will
not be allowed to weaken.
f)Adjustment of output amplitude: 0-16 levels.
g)The timer of therapeutic equipment: 15 minutes.
h)Safety classification: internal electric source class and type BF equipment
i)Boundary dimension:
0DLQXQLW0RGHO.75ĭ
j)Product software version No.: A/0
1.The device is convenient to carry, and is suitable for personal health care;
2.
Electrode patch intelligent detection and protection, users use more secure;
3.Electric pulse combination, 16 level strength can be adjusted, according
to personal preferences to adjust the need;
4.Integrated fuselage design, functional operation, simple and clear;
5.Rechargeable lithium design, built-in power protection circuit, easy to use,
safe;
6.Contains a variety of massage procedures, can meet different massage
needs, suitable for a wider range of people;
(1)Preparation before use
Use the electrode plate for the first time, install the electrode plate to the
master machine of the massage machine, and then align the metal buckle
of the main machine to remove the film on the electrode sheet;The
electrode film is affixed to the main body of the massage instrument;Attach
the main body to the use part;
(2)Switch machine
Press key, the host boot starts, at the same time will send the “drop”
sound prompt, the strength output default is 0, the host starts to work; the
default is the first mode of operation. Press the key again, then turn off
the machine.
(3)Charge
If you can't turn on the machine, or turn it off automatically, or when the
buzzer is alerted for 30s, it indicates that the product needs to change the
battery.
(4)Regulating intensity
Users can press the “+” key, set the strength, press the “-” key, you can
reduce the strength; short press the “+” key or “+” key strength will increase
k)Service life: the shelf life of subject device is 2 years; the use life of the
(OHFWURGH3DGLVWLPHVDQGVKHOIOLIHLV\HDUV
7.3 Product Environmental Requirement
a)Normal work environmental requirements:
Environment temperature: +5~+40;
(QYLURQPHQWKXPLGLW\5+
$WPRVSKHULFHQYLURQPHQWFRQGLWLRQVK3DK3D
b)Storage environment requirements:
Environment temperature: -25~+55;
Environment humidity: 0-93%RH;
Atmospheric environment conditions: 500hPa-1060hPa.
c)Transport environment requirements:
Environment temperature: -25~+55;
Environment humidity: 0-93%RH;
Atmospheric environment conditions: 500hPa-1060hPa.
ⒺⒾ
ⒺⒿ