32 ADB_ADJ-BA-e-1830
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other objects.
Electromagnetic fields / static charging (choose a
different location / switch off interfering device if
possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose a
different location / switch off interfering device if
possible)
14.1 Error messages
E Weighing range exceeded; the set load exceeds the weighing range.
Unload the balance.
E Weighing range exceeded, e.g. weighing plate not installed.
CAL-no Adjustment error.