Balance is for determining weighing value of material, manually placed, read after stable value.
Avoid dynamic weighing, permanent load, impact shock, overloading, hazardous locations, structural alterations.
Guarantee invalid for non-observation of guidelines, use outside described applications, alterations, damage, wear, improper installation.
Metrology features and adjusting weights must be checked regularly for quality assurance.
Read operating instructions carefully before erecting and commissioning, even with prior experience.
Device may only be operated and looked after by trained staff members.
Check packaging and device for visible signs of external damage upon delivery.
Keep original packaging for returns, use only original packaging, secure parts against shifting.
Select correct location for quick and precise work, ensure firm/level surface, avoid heat, draughts, jarring, humidity, dust.
Carefully remove balance from packaging, remove plastic wrap, position, install horizontally, list accessories.
Connect batteries or mains supply. Balance conserves power via auto-off. Use original KERN power circuits.
A warm-up time of 3 minutes stabilizes measured values after switching on.
Coordinate balance to local gravity. Perform adjustment during start-up, location change, or temperature variation.
Weigh objects too large/shaped for scale using underfloor weighing. Hang item on hook below the balance.
Details the function of the ON/TARE and OFF keys for brief presses and 10-second presses.
Switch on balance, perform auto-diagnostics, display '0'. Indicates 'E' for overload. Switch off with OFF key.
Place tare cup, press TARE to zero. Fill object, read value. Repeatable for components. Total weight shown as negative.
Change units by pressing and holding the ON/TARE key. Displays available units and conversion factors.
Auto-Zero function tares zero changes. Activate/deactivate Zero-Tracking via key combination.
Automatic switch off function for battery operation can be activated/deactivated in the menu.
Allows factory setting reset by pressing specific key combinations. Confirms with OFF key.
Disconnect power before cleaning. Use mild suds, soft cloth. Remove residue carefully. Clean spills immediately.
Device may only be opened by trained, authorized service engineers. Disconnect from mains before opening.
Dispose of packaging and device in compliance with valid national or regional law.
Possible causes: balance not switched on, empty/wrongly inserted batteries, or power supply interruption.
Causes: draught/air movement, table/floor vibrations, or foreign matter on weighing plate.
Causes: display not zeroed, incorrect adjustment, temperature fluctuations, or electromagnetic fields.