EasyManuals Logo
Home>KERN>Scales>KB 10000-1

KERN KB 10000-1 User Manual

KERN KB 10000-1
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
572/573/KB/DS/FKB/FCB/PCB-BA-e-0853
21
7.4 RS 232 C Data output via interface RS 232 C
RS 232 C Data output via interface RS 232 C
General information
As a condition for the data transfer between the balance and a peripheral device (for instance printer, PC ...)
both devise have to be set on the same interface parameter (for instance baud rate, parity ...).
7.4.1 There are 4 methods for the data output via RS 232 C
Data output via PRINT-Key
The printing process can be released by the PRINT-key. In this case the settings AUTOPRINT and AUT-
PRINT PC should be deselected.
AUTOPRINT (Data output, after having loaded the balance)
The setting AUTOPRINT is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT is active the actual weighing value will be sent via the RS 232 interface when the balance has
been unloaded and then loaded after having achieved the stability.
AUTOPRINT PC (Continuous data output)
The setting AUTPRINT PC is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT PC is active the actual weighing values will be sent continuously via the RS 232 interface.
Data output by transfer of remote controls
The following functions can be released by the remote controls that will be transferred as ASCII signs to the
balance.
t Tare.
w a weighing value (or unstable) is sent via RS 232 interface.
s a stable weighing value is sent via RS 232 interface.
If the balance receives the command w or s, it acts without printing delay.
7.4.2 Description of the data transfer
Structure of each data transfer:
Without Numerator:
Bit-Nr. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
B B B B B B B B 0
.
0 B g B B CR LF
B*
B*: = Blank or % in the range of zero point.
B, 0,
.
, g: = Blank or weighing value with unit, depending on the load on the weighing plate.
CR: = Carriage Return
LF: = Line Feed
With Enumerator:
Bit.Nr. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
N N N B* B B B B B B B 0
.
0 B G B B CR LF
N: = Enumerator
7.4.3 Numerator
The Numerator is found under the menu point “Printer” and can be activated and deactivated. When editing
data with the use of the print option, the range of the numerator will increase by one place.