2.6 KERN PCB
KERN PCB 200-3B PCB 240-3B PCB 2000-2B PCB 2400-2B PCB 10K0.05B
Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g 0.05 g
Weighing range (max) 201 g 241 g 2010 g 2410 g 10.100 g
Taring range (subtrac-
tive)
201 g 241 g 2010 g 2410 g 10.100 g
Reproducibility 0.001 g 0.001 g 0.01 g 0.01 g 0.05 g
Linearity ± 0.003 g 0.003 g 0.03 g 0.03 g 0.15 g
Smallest piece weight 0.001 g 0.001 g 0.01 g 0.01 g 0.05 g
Recomm. adjusting
weight F1 (not sup-
plied)
200 g 200 g 2000 g 2000 g 10 kg
Humidity of air
max. 80% rel. (not condensing)
Appropriate for verifica-
tion
no
Stabilization time
(typical)
3 sec.
Permitted environ-
mental temperature
+ 10 °C ... + 40 °C
Vibration filter yes
Housing
(B x D x H) mm
170 x 250 x 85
Weighing plate,
stainless steel
Ø 81 Ø 81 130 x 130 130 x 130 150 x 170
Units
see menu
Vibration filter
yes
Weight kg (net)
1 1 1.5 1.5 1.7
Data interface yes (RS232)