4 Move down to the parameters” level.
The multimeter will wait for the security code to be entered.
^000000 CODE
5 Unsecure the multimeter by entering the security code.
The security code is set to “
HP034401 when the multimeter is shipped
from the factory. The security code is stored in non-volatile memory, and
does not change when power has been off or after a remote interface reset.
Use
< and > to move left or right between digits. Use
or
to increment or decrement numbers.
If you have not changed the security code from its factory setting, you
can unsecure the multimeter by entering “034401” from the front panel.
Only the last six characters are recognized from the front panel and
they must be numeric characters only.
^034401 CODE
We recommend that you use a unique SECURE code for each multimeter
to obtain the maximum benefit from the electronic calibration security
features of the multimeter.
To secure the multimeter again, return to the parameter level of the
UNSECURED command and enter a new security code.
For further information on the calibration security features of the
multimeter, seeCalibration Security Code” on page 73.
4 3 0
1 0 4
Auto/Man
ENTER
Chapter 3 Menu Tutorial
A Front-Panel Menu Tutorial
54

Table of Contents

Related product manuals