Service and Maintenance 1
Keysight E36200 Series Service Guide 17
User Replaceable Parts
You can find the instrument support part list from the Keysight's Test &
Measurement Parts Catalog at http://www.keysight.com/find/parts.
Battery Replacement
The internal battery powers the real-time clock. The primary function of the clock
it to provide a time stamp for the internal file system. If the battery fails, the clock
and time stamp function will not be available. No other instrument functions are
affected.
Under normal use at room temperature, the lithium battery has a life expectancy
between seven and ten years. Note that battery life will be reduced if the
instrument is stored for a prolonged period at temperatures above 40 degrees
Celsius.
The battery type is CR 2032 (3V).
WARNING
SHOCKHAZARD
Only qualified, service-trained personnel who are aware of the hazards
involved should remove instrument covers. Always disconnect the power
cable and any external circuits before removing the instrument cover.Some
circuits are active and have power for a short time even when the power
switch is turned off.