Verification and Adjustments 2
Keysight E36200 Series Service Guide 77
Constant current tests
Current programming
Zero current output (I
0
)
Max V, 0 A –0.003 A ____________ 0.003 A
Maximum current ouput (I
max
)
Max V, 10 A 9.987 A ____________ 10.013 A
Current readback
Zero current measured over interface Max V, 0 A
I
0
– 0.003 A
____________
I
0
+ 0.003 A
Maximum current measured over interface Max V, 10 A
I
max
– 0.013 A
____________
I
max
+ 0.013 A
CC load regulation (I
load
- I
short
)
Max V, 3.333 A –0.5833 mA ____________ 0.5833 mA
Max V, 10 A –1.250 mA 1.250 mA
CC line regulation (I
lowline
- I
highline
)
Max V, 3.333 A –0.5833 mA ____________ 0.5833 mA
Max V, 10 A –1.250 mA ____________ 1.250 mA
Low range current measurement
Maximum low range current measured over interface 60 V, 0.05 A
I
low max
– 205 μA
____________
I
low max
+ 205 μA
Table 2-12 Test record form - Keysight E36232A (continued)
E36232A Report number _______________ Date _______________
Description DUT setting Lower limit Result Upper limit

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E36234A and is the answer not in the manual?

Keysight Technologies E36234A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE36234A
CategoryPower Supply
LanguageEnglish

Related product manuals