Verification and Adjustments 2
Keysight E36200 Series Service Guide 81
Constant current tests
Current programming
Zero current output (I
0
)
Max V, 0 A –0.003 A ____________ ____________ 0.003 A
Maximum current ouput (I
max
)
Max V, 10 A 9.987 A ____________ ____________ 10.013 A
Current readback
Zero current measured over interface Max V, 0 A
I
0
– 0.003 A
____________ ____________
I
0
+ 0.003 A
Maximum current measured over
interface
Max V, 10 A
I
max
– 0.013 A
____________ ____________
I
max
+ 0.013 A
CC load regulation (I
load
- I
short
)
Max V, 3.333 A –0.5833 mA ____________ ____________ 0.5833 mA
Max V, 10 A –1.250 mA 1.250 mA
CC line regulation (I
lowline
- I
highline
)
Max V, 3.333 A –0.5833 mA ____________ ____________ 0.5833 mA
Max V, 10 A –1.250 mA ____________ ____________ 1.250 mA
Low range current measurement
Maximum low range current measured
over interface
60 V, 0.05 A
I
low max
– 205 μA
____________ ____________
I
low max
+ 205 μA
Table 2-14 Test record form - Keysight E36234A (continued)
E36234A Report number _______________ Date _______________
Description DUT setting Lower limit
Result
Upper limit
Output 1 Output 2