3 Remote Interface Reference
108 Keysight E364xA User’s and Service Guide
The CURR:STEP? DEF returns the minimum resolution of your instrument. The
immediate current level increases or decreases by the value of the step size.
For example, the output current will increase or decrease 10 mA if the step size
is 0.01. At *RST, the step size is the value of the minimum resolution.
CURRent:STEP? {DEFault}
Return the value of the step size currently specified. The returned parameter is
a numeric value. DEFault gives the minimum resolution of the step size in unit
of amperes.
CURRent:TRIGgered {<current>| MINimum|MAXimum}
Program the pending triggered current level. The pending triggered current
level is a stored value that is transferred to the output terminals when a trigger
occurs. A pending triggered level is not affected by subsequent CURRent
commands.
CURRent:TRIGgered? [MINimum|MAXimum]
Query the triggered current level presently programmed. If no triggered level is
programmed, the CURRent level is returned.
Example
The following program segments show how to use the CURR UP or CURR DOWN
command to increase or decrease the output current with the CURR:STEP
command.
CURR:STEP 0.01
Set the step size to 0.01 A
CURR UP
Increase the output current
CURR:STEP 0.02
Set the step size to 0.02 A
CURR DOWN
Decrease the output current

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3645A and is the answer not in the manual?

Keysight Technologies E3645A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3645A
CategoryPower Supply
LanguageEnglish

Related product manuals