Service and Maintenance 8
E364xA User’s and Service Guide 231
11 A high current calibration point is displayed. Enter the computed value
(DVM reading divided by the shunt resistance) by using the knob and
resolution keys. Press the Calibrate key to save the changes.
AHigh current
calibration
12 Turn off the Power switch to exit the calibration menu. Exit CAL MODE
Table 8-9 Calibration record