EasyManua.ls Logo

Keysight Technologies FieldFox N9914A

Keysight Technologies FieldFox N9914A
179 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
120 Keysight FieldFox Handheld Analyzers Service Guide
Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
7-
Replacing the Strap Handle
1. Separate the front and rear cases to the positions shown in Figure 7-7. See
Separating the Front and Rear Cases.
2. Note the orientation of the strap handle, then lift it off of the dowel pins at
each end.
3. Slip the new strap handle over the dowel pins.
4. Reinstall the rear case onto the front case. See Combining the Front and
Rear Cases.
Replacing the I/O Side Panel, Doors, and Gasket
1. Separate the front and rear cases. See “Separating the Front and Rear
Cases” on page 115.
2. Refer to Figure 7-10. Disconnect the A5 System board to A3 FPIB cable,
W14, at the A3 FPIB connector. If replacing this cable, disconnect both
ends.

Table of Contents

Other manuals for Keysight Technologies FieldFox N9914A

Related product manuals