Keysight FieldFox Handheld Analyzers Service Guide 33
Tests and Adjustments
Before Returning an Instrument to a Customer
Before Returning an Instrument to a Customer
When finished servicing an FieldFox analyzer, perform the following before
returning the unit to the customer:
Set the Display Colors to Indoor and the display Brightness to 80% via
System (7) > Preferences > Preferences, etc.
Perform an Operational Check as described in the following section.
Connect and disconnect the external battery charger while the FieldFox is
operating and confirm that the power cord icon is displayed appropriately
at the top of the display.
Charge the battery to 80% or higher.
Connect FieldFox to the internet.
—Press System > System Configuration > More > update time using
Internet.