EasyManua.ls Logo

Keysight Technologies N2791A - Offset Zero Calibration Procedure

Keysight Technologies N2791A
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36 Keysight N2791A/N2891A High-Voltage Differential Probes User Guide
Offset Zero Calibration Procedure
Offset Zero Calibration Procedure
The N2791A/N2891A differential probe can be adjusted for zeroing out the
probe’s offset voltage using the trimmer tool supplied with the probe.
Follow this procedure to perform the offset zero calibration.
1 Connect the N2791A/N2891A probe to channel 1 of the scope. Turn on
the probe power. You may use the USB power code or batteries to
power the probe. Select the probe attenuation ratio to 10:1 (for
N2791A) or 100:1 (for N2891A).
2 Short the + and - probe inputs together with the hook tips.
3 Turn on power to the oscilloscope and the probe, and leave them on for
> 30 minutes so they can stabilize.
4Press [Default Setup] and [Auto Scale] on the oscilloscope.
5 Press the channel 1 button, then press the Probe softkey, and then set
the units to Volt and the probe attenuation to 10:1 (for N2791A) or
100:1 (for N2891A).
6 Set the scope to DC coupled mode and the scope offset to 0 Volt.
7 Set the oscilloscope to Averaging mode (x8 or higher) or
high-resolution mode to reduce oscilloscope noise.
8 Using the trimmer tool that comes with the probe, adjust the probe
offset voltage to 0 volts
.