Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 436
Functional Tests
Leveled Output Power
c. If using a power meter, configure the power meter as follows:
1. On the power meter, press the Frequency Cal Fac button.
2. If applicable, select a power meter channel.
3. Use the arrow keys to enter the frequency at which to
measure the power.
d. Measure the output power level.
e. Repeat steps a through d for the remaining frequencies in the table,
and confirm that the power level at each point is within limits.
With Modulation 13.Preset the signal generator: Press Preset.
14.Select the factory-supplied waveform SINE_TEST_WFM:
a. Press Mode > Dual ARB > Select Waveform.
b. Highlight the SINE_TEST_WFM waveform.
c. Press Select Waveform.
Table 17-3 Output Power Without Modulation
Output Power Without Modulation
Frequency (MHz) Amplitude (dBm) Limits (dB)
50 7 ±0.6
100 7 ±0.6
250 7 ±0.6
338 7 ±0.6
425 7 ±0.6
538 7 ±0.6
675 7 ±0.6
850 7 ±0.6
1075 7 ±0.6
1350 7 ±0.6
1700 7 ±0.6
2150 7 ±0.6
2700 7 ±0.6
3400 7 ±0.6
4300 7 ±0.6
5400 7 ±0.6