EasyManua.ls Logo

Keysight Technologies Truevolt Series - Page 14

Keysight Technologies Truevolt Series
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Declassification and Security Product Name: Truevolt Series of Digital Multimeters
Model Numbers: 34460A, 34461A, 34465A, 34470A
Page 14 of 17
The LAN port provides the following services, which can be selectively disabled:
a) LAN
b) VXI-11
c) Sockets
d) Telnet
e) Web
f) mDNS
g) HiSLIP
To disable LAN services:
On the front panel, press:
[Utility] > I/O Config > LAN Off/On, or use
[Utility] > I/O Config > LAN Settings > LAN Services, then select the
interface to enable or disable
On the remote interface, use:
SYSTem:COMMunicate:ENABle {ON|OFF}, <interface>
For mDNS on the remote interface, use:
LXI:MDNS:ENABle {ON|OFF}
To disable USB:
On the front panel, press:
[Utility] > I/O Config > USB Settings > USB SCPI Off/On
On the remote interface, use:
SYSTem:COMMunicate:ENABle {ON|OFF}, USB
To disable GPIB:
On the front panel, press:
[Utility] > I/O Config > GPIB Settings > GPIB Off/On
On the remote interface, use:
SYSTem:COMMunicate:ENABle {ON|OFF}, GPIB
USB Mass Storage Device Security
Not Supported
Controlling the USB device MTP (driverless) connection service

Related product manuals