Description of Parts
12
  Zero Adjustment Dial (ZERO ADJ)
Use the zero adjustment dial to correct for the effect of a voltage 
offset or temperature drift on the unit. The probe should be always 
be zeroed after demagnetization. Refer to “Demagnetization and 
Zero Adjustment” on page  16.
  Coarse Adjustment Trimmer
This adjustment should only be carried out if the probe offset is 
outside the range of the zero adjustment dial.
The output of this unit is terminated internally. The high-impedance such as 
1 MΩ input impedance will be automatically selected on the Infiniium, as 
the Terminator is plugged into the AutoProbe interface. With an input 
impedance of 50 
Ω, accurate measurement is not possible.
The probe output sensitivity 0.1 V/A will be automatically selected, as the 
Terminator is plugged into the AutoProbe interface. With an different 
output sensitivity, accurate measurement is not possible.