Keysight E8257D/67D & E8663D PSG Signal Generators Installation Guide 15
 Operation Verification 
Checking the Maximum Leveled Power 
If the signal generator is not equipped with Option 007, press Sweep /List > Configure Step Sweep.
7. Press  Freq Start and enter the start frequency value listed in Table 3-1,  Table 3-2, or Table 3-3 that 
corresponds to the amplitude value entered in step 5.
8. Press  Freq Stop and enter the corresponding stop frequency listed in Table 3-1,  Table 3-2, or 
Table 3-3.
9. Press  # Points and enter the number of corresponding points listed in Table 3-1,  Table 3-2 or 
Table 3-3.
10. Press  Return > Sweep > Freq. The SWEEP annunciator is now on and a continuous sweep will begin. 
During the sweep, the progress bar becomes active on the display, indicating the status of the 
current sweep.
11. Watch the display and ensure that the Unlevel annunciator does not appear at any time during 
a full sweep. 
The  Unlevel annunciator may turn on and off, or remain on; both conditions indicate a failure. 
If this occurs, refer to “Maximum Leveled Power Check Failure” on page 19 for further instructions.
12. Press  Sweep > Off.
13. Repeat steps 5 through 12 for each frequency range specific to your instrument model number 
and options.