Keysight E8257D/67D & E8663D PSG Signal Generators Installation Guide 27
Operation Verification
Checking the Output Power
Table 3-9 Frequency and Power Level Limits (For PSGs with s/n prefixes >=US5042/MY5042)
a
Instrument
Model
Frequency
Amplitude
Standard Option 1EU Option 1E1 Options 1EU
and 1E1
E8257D
Option 520
250 kHz 13 dBm / ±0.6 16 dBm / ±0.6 11 dBm / ±0.6 15 dBm / ±0.6
3.2 GHz 13 dBm / ±0.8 16 dBm / ±0.8 11 dBm / ±0.8 15 dBm / ±0.8
20 GHz 13 dBm / ±0.8 20 dBm / ±0.8 11 dBm / ±0.8 18 dBm / ±0.8
E8257D
Option 521
250 MHz N/A N/A 19 dBm / ±1.3 N/A
1 GHz N/A N/A 24 dBm / ±1.0 N/A
6 GHz N/A N/A 28 dBm / ±1.0 N/A
20 GHz N/A N/A 27 dBm / ±1.0 N/A
E8257D
Option 532
and
Option 540
250 kHz 9 dBm / ±0.6 15 dBm / ±0.6 7 dBm / ±0.6 14 dBm / ±0.6
3.2 GHz 9 dBm / ±0.8 15 dBm / ±0.8 7 dBm / ±0.8 14 dBm / ±0.8
20 GHz 9 dBm / ±0.8 18 dBm / ±0.8 7 dBm / ±0.8 16 dBm / ±0.8
40 GHz 9 dBm / ±0.9 14 dBm / ±1.0 7 dBm / ±0.9 12 dBm / ±1.0
E8257D
Option 550
and
Option 567
250 kHz 5 dBm / ±0.6 14 dBm / ±0.6 3 dBm / ±0.6 13 dBm / ±0.6
3.2 GHz 5 dBm / ±0.8 14 dBm / ±0.8 3 dBm / ±0.8 13 dBm / ±0.8
20 GHz 5 dBm / ±0.8 14 dBm / ±0.8 3 dBm / ±0.8 13 dBm / ±0.8
40 GHz 5 dBm / ±0.9 11 dBm / ±1.0 3 dBm / ±0.9 9 dBm / ±0.9
50 GHz 5 dBm / ±1.3 9 dBm / ±1.3 3 dBm / ±1.3 9 dBm / ±1.3
60 GHz 5 dBm / ±1.5 9 dBm / ±1.5 3 dBm / ±1.5 9 dBm / ±1.5
67 GHz 5 dBm / ±1.5 9 dBm / ±1.5 3 dBm / ±1.5 8 dBm / ±1.5
E8267D
Option 520
250 kHz 13 dBm / ±0.6 N/A N/A N/A
3.2 GHz 13 dBm / ±0.8 N/A N/A N/A
20 GHz 18 dBm / ±0.8 N/A N/A N/A
E8267D
Option 532
and
Option 544
250 kHz 12 dBm / ±0.6 N/A N/A N/A
3.2 GHz 12 dBm / ±0.8 N/A N/A N/A
20 GHz 14 dBm / ±0.8 N/A N/A N/A
40 GHz 12 dBm / ±1.0 N/A N/A N/A
44 GHz 10 dBm / ±0.9 N/A N/A N/A
E8663D 250 kHz 13 dBm / ±0.6 16 dBm / ±0.6 11 dBm / ±0.6 15 dBm / ±0.6
3.2 GHz 13 dBm / ±0.8 16 dBm / ±0.8 11 dBm / ±0.8 15 dBm / ±0.8
9 GHz 13 dBm / ±0.8 20 dBm / ±0.8 11 dBm / ±0.8 18 dBm / ±0.8
a
For instrument with Type-N connectors (Option 1ED), specifications are degraded typically 0.2dB above 18GHz