Introduces the purpose and scope of the ECal module reference guide.
Assumes user training in connector care for accurate calibration.
Explains how adapters, ECal modules, and gages are referenced by connector type.
Compares ECal accuracy versus mechanical calibration, highlighting benefits.
Discusses factors affecting calibration validity and when to recalibrate.
Details model numbers, connector types, and frequency ranges for ECal modules.
Details N755xA series RF two-port ECal modules, their types, and options.
Details N4431D series RF four-port ECal modules, their types, and options.
Details N469xD series microwave two-port ECal modules, their types, and options.
Details N4432D/N4433D series microwave four-port ECal modules, types, and options.
Lists network analyzer models compatible with ECal modules.
Verifies the ECal kit case and contents are undamaged and complete.
Explains how to locate and identify ECal kit and module serial numbers.
Lists the standard items included in various ECal kits.
Describes how options customize ECal modules for connector types and features.
Provides guidelines for safe and effective handling and operation of ECal modules.
Details measures to protect against electrostatic discharge when handling ECal modules.
Guides on inspecting and maintaining connectors to ensure longevity and performance.
Specifies maximum input power and DC levels to avoid damaging ECal modules.
Outlines the temperature and humidity ranges for optimal ECal module performance.
Provides steps for initiating a calibration procedure with an ECal module.
Explains the meaning of different LED colors and patterns on the ECal module.
Details step-by-step procedures for performing calibrations on various analyzers.
Specific instructions for calibrating with PNA and PXIe network analyzers.
Specific instructions for calibrating with ENA network analyzers.
Specific instructions for calibrating with FieldFox handheld analyzers.
Lists supported calibration types and advanced settings for different analyzers.
Explains techniques for calibrating devices with non-connectable ports.
Directs users to resources for detailed information on performing ECal calibrations.
Describes methods to verify the accuracy and functionality of a completed calibration.
Basic validation method comparing measured data to factory data.
Uses verification kits for traceable accuracy validation of the measurement system.
Provides solutions for common issues based on ECal module LED status and possible causes.
Essential measures to protect against static electricity when handling sensitive components.
Guidelines for inspecting connectors for wear, damage, and contamination.
Detailed methods for cleaning coaxial and 7 mm connectors to maintain performance.
Procedure for cleaning coaxial connectors using compressed air or nitrogen.
Procedure for cleaning 7 mm connectors, including collet handling.
Explains the purpose, accuracy, and procedures for using connector pin depth gages.
Discusses the accuracy limitations and factors affecting connector gage measurements.
Step-by-step instructions for performing pin depth measurements using connector gages.
Essential techniques for making accurate and reliable connections to ECal modules.
Specific steps for connecting standard RF/Microwave connectors to ECal modules.
Specific steps for handling and connecting 7 mm connectors, including collet removal.
Proper use of torque wrenches to ensure correct connection tightness and prevent damage.
Recommendations for safely storing and handling ECal modules and accessories.
Information on the optional ruggedized storage box for ECal modules and accessories.
Defines key terms used in ECal module specifications for clarity.
Details operating temperature, humidity, and storage conditions for ECal modules.
Provides warranted electrical performance parameters for various ECal module series.
Detailed electrical performance data for the N755xA ECal module series.
Detailed electrical performance data for the N4431D ECal module series.
Detailed electrical performance data for N4432D and N4433D ECal modules.
Detailed electrical performance data for the N469xD ECal module series.
Lists non-warranted performance characteristics for ECal modules.
Describes mechanical properties like pin depth and connector details.
Explains pin depth, protrusion, and recession concepts for connectors.
Specific details on the mechanical characteristics of Type-N connectors.
Provides tables of typical pin depths, uncertainties, and limits for various connectors.
Information on downloading and using Keysight's calculator for measurement uncertainty.
Lists part numbers for Type-N ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 1.85 mm ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 2.4 mm ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 2.92 mm ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 3.5 mm ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 4.3-10 ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 7 mm ECal modules, adapters, caps, gages, and other items.
Lists part numbers for 7-16 ECal modules, adapters, caps, gages, and other items.
Lists general ECal accessories like cleaning supplies, ESD protection, and storage boxes.