10 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-350 Queue overflow
This is a specific code entered into the queue in lieu of the code
that caused the error. This message indicates that there is no more
room in the queue and an error occurred but was not recorded.
-330 Self-test failed; BBG Aux IO Test Failed: {name}
The Aux IO Loop back test has detected a failure. Contact the
service office for assistance.
-330 Self-test failed; BBG Feldspar output clock alignment failed: {name}
The BBG Feldspar output clock alignment did not successfully
complete. Contact the service office for assistance.
-330 Self-test failed; BBG {name} {name} Memory Test Failed: {name}
The BBG Arb waveform memory (SDRAM) test has detected a failure.
Contact the service office for assistance.
-330 Self-test failed; BBG{value}: {name}
A self-test executed on the baseband generator failed. Report this
error to service office.
-330 Self-test failed; The loop back test failed because the digital module
could not be set up.
To run a diagnostic loop back test, the instrument needs to
communicate with and set up the module. This error is reported when
the instrument cannot communicate with the module.