Keysight X-Series Signal Generators Error Messages 19
Error Messages
List of Error Messages
-313 Calibration memory lost; Cannot store cal: {name}err={value}
Cannot store specified cal because of a file error.
-313 Calibration memory lost; Error occurred while trying to load data from
previous internal IQ calibration.
An expected file was not found while trying to load internal IQ
calibration data.
-313 Calibration memory lost; Previous Internal IQ calibration data is
invalid. Data has been removed. A new user IQ calibration is
recommended.
An expected file was not found while trying to load internal IQ
calibration data. All other data has been removed. A new cal should
be performed.
-313 Calibration memory lost; Stored calibration data is invalid: {name}
An unexpected error occur while attempting to restore cal data. The
stored data is not valid. Service is required. Report this error to
service office.
-310 System error
*** No documentation ***
-310 System error; A SCPI driver failed to initialize ({name})
The hardware or driver for a particular SCPI parser has failed and
that communication path is not active. The hardware device is