20 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
specified in the error message. If this error persists, please
contact the service office.
-310 System error; A SCPI parser has unexpectedly exited.
A particular SCPI parser has failed and that communication path is
no longer active. If this error persists, please contact the
service office.
-310 System error; The ADC {name} has timed out during a conversion.
The returned data is not valid. Report this error to service
office.
-310 System error; Arb generation failed. File System failed to access
required file.
There was a problem opening, reading or writing to the specified
file.
-310 System error; Automatic reboot performed due to {name}
The instrument has rebooted itself because of an exception event
(detailed in the {name}). This is a internal firmware error. If this
error recurs after updating to the most recent firmware, report the
contents of this error message to the nearest sales and service
office.
-310 System error; Cal in use, cannot delete: {name}
Calibration file cannot be deleted since it is currently in use.
Error caused by diagnostic command.