Keysight X-Series Signal Generators Error Messages 45
Error Messages
List of Error Messages
-224 Illegal parameter value; Illegal parallel bus address
Internal error. The instrument tried to access an illegal parallel
bus address.
-224 Illegal parameter value; Insufficient privilleges to perform write
({name}).
The diagnostic write to an item has failed due to a lack of
permission.
-224 Illegal parameter value; Latch name not present
The latch name that was selected is not a valid latch in the
system.
-224 Illegal parameter value; Option 416 does not have a valid license.
*** No documentation ***
-224 Illegal parameter value; The OSR must be an even value.
RADio:ARB:GENerate:SINE generates a valid sine waveform and the
OSR must be an even value to generate a waveform with an even
number of samples.
-224 Illegal parameter value; The specified option cannot be enabled.
An error occurred when the user attempted to enable an option that
is not configured to be enabled.